Evaluación de desempeño del micropet focus 120 con los radionúclidos $^{11}$C, $^{13}$N y $^{18}$F

F. Trejo Ballado, N. Ortega López, R. Ojeda Flores, M.A. Ávila Rodríguez

Abstract


The aim of this work is to evaluate the characteristics involved in the data acquisition and image reconstruction of a microPET Focus 120 system. Energy resolution, spatial resolution, sensitivity, count rate and image quality characteristics were evaluated using different phantoms with three conventional positron emitter radionuclides used in PET imaging ($^{11}$C, $^{13}$N and $^{18}$F). Some tests performed in this work are from the new protocol NEMA NU4 -- 2008.

Keywords


MicroPET; performance evaluation; focus 120; image reconstruction